11. Calibrations for Energy vs Channel, NaI(Tl) Detectors

We are now ready to perform a sodium iodide energy calibration. The Energy Calibration program is reached by selecting the “Calibration” option from the Main Menu, see Figure 11-1, below, which shows the main menu. Please remember that this menu is the same for all “HELGE” software.

Chap 11 Mainmn4
Figure 11-1, This is the Main Menu with “Calibration” selected.

There are six steps which must be performed to assure that the calibration data are obtained properly:

  • Part 1, Verification of Dial Settings,
  • Part 2, Calibration of Detectors 1 Through “n,”
  • Part 3, Recording the Data,
  • Part 4, Obtaining a New Background,
  • Part 5, Statistical Errors in Channel Center Measurements, and
  • Part 6, Determining the “Dial Factors.”
The last step, “Part 6, Determining the Dial Factors,” really should be called “Part 1” because these factors must be determined before “semi-automatic” calibration work can actually begin. This information has been placed at the end of this chapter by the Editor’s choice since once the dial factors have been determined for a specific amplifier and detector, they seldom need to be changed. 

The menu which you receive is a function of the type of hardware being used. Figures 11-2 through Figure 11-4 show the menus from:

     the “Do-It-Yourself Whole Body Counter,”
     the “Quicky I” in vivo Counter,”
     the "BRC" Waste Counter.

The screens for the Quicky I, Quicky III, and Quicky VI are identical because they do not have any moving parts, i.e., there is no "Speed" calibration.

These three screens are shown below.

CALMNUDY.JPG
Figure 11-2, This is the Do-It-Yourself Calibration” menu with “Energy vs Channel” selected.

CALMNUQK.IMG
Figure 11-3, This is the "Calibration" menu for the "Quicky I," the "Quicky III," or the "Quicky VI" in vivo counters with the option "Energy vs Channel" selected. They are all the same because none of these have moving detectors, beds or turntables.

CALMNUDY.IMG
Figure 11-4, This is the "BRC Calibration” menu with “Energy vs Channel” selected.
 

The "Energy vs. Channel" program may be protected by demanding a password before allowing the technician to use the program as shown  in Figure 11-5 below.

CALMNUPW.IMG
Figure 11-5, The "Energy versus Channel" option may be protected by demanding a password before allowing the technician to use the program.
 

11.1. Energy Calibration, Part 1, Verification of Dial Settings

Upon entering the sodium iodide energy calibration routine, you are presented with the screen shown below. Examine it. It shows that we are looking for the 241-americium photopeak of 59.6 keV in channel 14.282. Likewise, we are looking for a high energy photopeak from 60-cobalt at 1173.2 keV, which should be found in channel 230.370. Remember that we are accounting for the non-linear response of NaI(Tl) detectors in determining the center channel values. Please refer to Chapter 10 for a complete discussion of the non-linear features of NaI(Tl).

Each of the preamplifiers has a dial position recorded in the parameters file. The computer prints these values and asks if the values are correct. You should inspect each of the preamplifiers to see whether or not the dial is actually set at these numbers. You have three choices. If the dials are all set correctly you can answer “YES” and the program will proceed.

If the dials are not set at these settings you have the other two choices:

  • If the dials are close to these setting, such as within 0.5 turns of these settings, you could actually put the dials at these setting, in which case you could answer “YES.”
  • You could correct the dial settings in the more proper method by typing “NO” and following the instructions given by the computer.
In this example we have found the dial positions to be correct, so we may proceed.
ECAL001.IMG
Figure 11-6, Verification of the Dial Settings.

Before you respond “Yes” to this screen, be sure that the energy versus channel calibration sources are in front of the detector(s) at the reference position(s). Please refer to Chapter 10 for further information.
 

11.2. Energy Calibration, Part 2, Calibration of Detectors 1 Through “n.”

The more recent versions of the “HELGE PC” software (starting with Version 2.0, we are now on Version 5.0 for DOS and the w\Windows version is almost finished) allow multiple detectors to be calibrated simultaneously. The actual calibration is accomplished by placing check sources in front of each of the detectors, which you should have done before you reached this screen.  

Figure 11-7, below, shows the spectrum as it is being accumulated. In this particular example the total counting time was 60 seconds. We see that we still have 21 seconds to complete the data acquisition. The total sampling time comes from the parameters. See Figure 9-17.

Note the red markers. These are located in the channels specified in the “Examine/Change Parameters” menu, Option 8, as shown in Figure 9-17. Typically, the sources used are 241-americium and 60-cobalt. Therefore, the program will read the parameters value for the correct center channels and place markers in the integer value (rounded off to the nearest whole channel). Thus, in Figure 11-7 we see red markers in channels 14 and 230. The marker in channel 230 will give us an idea immediately of where the lower energy peak of 60-cobalt should fall. It is very important that we look at this graph. In the past there have been a number of instances where the operator has not paid attention to the position of the 60-cobalt photopeaks and, because of carelessness, has allowed the 1332.5 keV photopeak to fall in the channel where the 1173.2 keV photopeak should fall. Obviously, this would give poor results. By using the “Up Arrow” or “Down Arrow” you may look at each detector in a multi-detector system. Don’t forget to check each detector for the proper location of the photopeak with reference to the red marker.
 

ECAL002.IMG
Figure 11-7, Calibration of a four-detector system, data acquisition screen.

Now look at Figure 11-8, below. At the end of the counting time the computer prints the results. If this figure could be shown in color on the printed page, we would see that the lines for detectors 1 and 3 would be high-lighted in yellow against a red background. This is because both of these detector dial settings need adjustment. The 241-americium photopeak for detector 1 was found in channel 14.23 and the 60-cobalt photopeak was found in channel 233.64. This is 3.27 channels too high, as the data for detector 1 shows. The limits of deviation from the correct 241-americium and 60-cobalt photopeak center are found in the “Examine/Change Parameters” menu, Option 8, Figure 9-17, lines 8 and 9. The new dial setting has been calculated to be 4.84 compared to its original setting of 4.94 (see Figure 11-6). The calculation is very simple: divide the number of channels by which the photopeak center has deviated from the correct channel, -3.27 channels for 60-cobalt, by its dial factor, which was 32.4 in this example. Thus:

     _VP_EQN_0.GIF .

This is the change in dial setting, so if we add the original dial setting of 4.94 we obtain

     _VP_EQN_1.GIF

The same calculation is done for detector 3. The two lines at the bottom of the screen tell us that we must change the dial settings (for detectors 1 and 3, the ones highlighted in yellow) and press “F2” to calibrate again.
 

ECAL003.IMG
Figure 11-8, One of the steps in the calibration of energy versus channel.

Stated in outline form the computer does the following:

a. subtract the center channel from the correct center channel

      The computer found the center channel at 233.64 but the correct center channel is 230.37. Therefore, for this first step we have:

     _VP_EQN_2.GIF

b. divide the channel difference by the dial factor

      The dial factor is 32.4 channels per 1.0 revolutions of the dial. Therefore, the difference in dial setting is:

     _VP_EQN_3.GIF

c. add this quotient to the old dial setting to obtain the new dial setting:

Thus,       _VP_EQN_4.GIF

Figure 11-9, below, shows the next screen, the one obtained after we made the adjustments shown on the previous page. This is not a screen which you would expect to see frequently. The significant part of this screen is the fact that we have a different note at the bottom of the red portion:

  • “Low Peak not within limits. . Adjustment might be required.”
ECAL004.IMG
Figure 11-9, Calibration of Detector 1, third screen.

If you could see this screen in color, you would note that the only portion in yellow was the 241-americium center channel for detector 1 which shows a value of 13.74. This is lower by 0.54 channels than the desired center channel of 14.28 - the difference is 0.54 channels. If you look at Figure 9-17, line 8 (at least for this particular set of parameters) the acceptable error for the 241-americium center channel is 0.5. Therefore, this result is 0.04 channels too low. For this example we have deliberately made very careful adjustments to the ADC “Zero” control. This is something which an inexperienced technician should not do. The limit of deviation from the correct 241-americium photopeak center can be as much as 1.0 channels without seriously affecting the gain and zero shift.

(Editor’s note: We even hesitated to mention this adjustment because an inexperienced person could cause more trouble than an error of only 0.04 channels would cause. Please give serious consideration to doing nothing, even if you experience 241-americium center channel errors of as much as 0.50 channels.)

After we made the careful adjustment, we pressed “F2” and continued.
 

ECAL006.IMG
Figure 11-10, Calibration of Detector 1, fourth screen.

Figure 11-10, above, shows the next screen, the one obtained after we made the adjustments shown on the previous page. This time detectors 1 and 4 were highlighted in yellow, signifying that adjustments of the dial settings were necessary.

ECAL005.IMG
Figure 11-11, Calibration of Detector 1, fifth screen

After making them, we obtained the screen shown in Figure 11-11, above. Note that the background screen for the results is blue. This signifies that all detectors are within their stated parameters. The “Energy versus Channel” calibration is finished.
 

11.3. Calibration, Part 4, Recording the Data

As the calibrations were being performed, all important information was being transmitted to the printer. When the calibrations are complete, as shown at the bottom of Figure 11-11, above, and you must press the “ESC” key. This will instruct the program to save the new dial settings. The printer will complete the printing.

This information is valuable for your Quality Assurance records. It is also valuable for Helgeson Maintenance Personnel in monitoring the performance of the system. An example of this printed record is shown in Table 11-1, immediately below this paragraph, and in Tables 11-2 and 11-3, below.

NaI Energy Calibration
11/16/93  12:12:16
Calibration time (sec)        60 seconds
NaI Low energy peak     Am-241     59.537 keV    Channel     14.280
NaI High energy peak    Co-60    1173.208 keV    Channel    230.370
Pre-amp #1 set at 4.94
Pre-amp #2 set at 5.83
Pre-amp #3 set at 5.20
Pre-amp #4 set at 4.93

    DETECTOR #    LOW CHANNEL    HIGH CHANNEL    DELTA    NEW SETTING

             1                           14.234                    233.638                   -3.268                4.84
             2                           14.273                    230.107 
                   0.263                5.83     *
             3
                           14.257                    231.463                   -1.093                5.17    *
             4 
                          13.928                    230.953                   -0.583                4.93

Pre-amp #1 set at 4.84
Pre-amp #2 set at 5.83
Pre-amp #3 set at 5.17
Pre-amp #4 set at 4.93

New dial settings were NOT saved to disk

Calibration is incomplete

Table 11-1, Typical written calibration report.

Table 11-1 corresponds to the problem which we saw at the bottom of Figure 11-9. As we stated, this is an unusual condition but one which should be recorded in the Quality Assurance records.

The data in Table 11-2, below, is much more typical of the calibration record.  Note that only two iterations were necessary for the four detectors to be returned to their properly calibrated conditions.
 

NaI Energy Calibration
11/19/93  08:58:33 Calibration time (sec)        60 seconds
NaI Low energy peak     Am-241     59.537 keV    Channel       14.269
NaI High energy peak       Co-60  1173.200 keV    Channel    230.408
Pre-amp #1 set at 4.57
Pre-amp #2 set at 5.60
Pre-amp #3 set at 4.92
Pre-amp #4 set at 4.20

DETECTOR #    LOW CHANNEL    HIGH CHANNEL    DELTA     NEW SETTING
   
      1                          13.744                    299.980                    0.428                4.57
    
     2                          13.877                    230.834                   -0.426               5.60
    
                              14.170                    229.878                    0.530               4.92
    
     4                          13.776                    230.194                    0.214               4.20

    DETECTOR #    LOW CHANNEL    HIGH CHANNEL    DELTA     NEW SETTING

         1                              13.995                    230.223                   0.185              4.57
         2                              14.181                    231.169                  -0.761              5.60
         3                              14.584                    229.940                   0.468              4.92
         4                              14.040                    230.758                 -0.350              4.20

Pre-amp #1 set at 4.57
Pre-amp #2 set at 5.60
Pre-amp #3 set at 4.92
Pre-amp #4 set at 4.20

New dial settings were saved to disk

Calibration completed

Table 11-2, Typical written Calibration Report, Part 2.

11.4. Energy Calibration, Part 5, Obtaining a New Background

Now that the counter has been calibrated, you must obtain a new background. Remove the source from the counter and store it safely away from the counter. Select the “Quick Count” option of the Data Acquisition Program (Option 1 from the “Main Menu”), and obtain a normal background.
 

WARNING:
If the gain or zero was adjusted, you must wait until a new background has been collected before starting more whole body counts. Data obtained with an incorrect background may be difficult or impossible to analyze correctly.


 
 
 

11.5. Statistical Errors in Channel Center Measurements

We have already seen in Chapter 3 that statistical errors are an expected part of all nuclear counting. This is also true of the determination of the center channel of a photopeak. Each channel used for the determination of the center of the photopeak will have errors which are proportional to the square root of the number of counts observed in that channel. Please refer to Chapter 6 for a discussion of the method for finding the center of a photopeak. Without going further into a difficult statistical explanation of the errors to be found in the determinations of the center of a photopeak, it is obvious that the errors will be a function of the number of counts obtained in the measurement. Thus, if you are using a weak source and are counting for a short period of time, the determination of the center will contain relatively large errors. Therefore, you should count for a reasonably long time. As a rough rule of thumb you should obtain at least 1000 counts in the center channel because then the counting error in that channel will be 

     _VP_EQN_5.GIF.

 Table 11-4, below, shows data obtained from a series of 45-second counts of an 850 nCi 60-cobalt source located in front of Detector #1 of a “Quicky III.”


Statistical Fluctuation in the 60-cobalt 1.17 MeV photopeak
as a function of time.

    Time, Minutes    Center Channel

          1.00                      230.19

        2.02                      230.37

        3.04                       230.19

        4.05                       230.59

        5.07                       231.26

        6.09                       230.07

        7.11                      230.42

        8.13                      229.94

      10.16                      230.23

      11.18                      230.67

      12.20                      230.24

      13.22                      230.90

      14.23                      229.75

      16.27                      230.48

      17.29                      230.35

      18.31                      230.29

      19.32                      230.51

      20.34                      230.34

Table 11-4, This table shows the random statistical fluctuations in the 60-cobalt 1.17 MeV photopeak as a function of time.

Figure 11-12, below, shows a graph of the same data. Note that the fluctuation in the center channel is small but real. The reader is invited to make similar measurements of his own using different counting times to see the influence of time on the determination of the center channel. This may be done in “background” mode or you may obtain the “DRIFT” program from Helgeson Scientific Services which allows you to select the counting time, the delay between counts, and the number of tests to be made. The output is a table of data points for the various photopeaks found as well as the apparent resolution of these photopeaks.

DRIFT.IMG
Figure 11-12, Example of the Random Statistical Fluctuation of the 60-cobalt 1173.2 keV photopeak.

Figure 11-13, below, shows the long-term drift of an NaI(Tl) detector if it is continually subjected to a strong source of radioactivity. It will recover if the source is removed.
 
 

LONGDRFT.IMG
This shows the long-term drift which may occur if a strong source is left in front of the NaI(Tl) detector for many hours.

11.6. Energy Calibration,Part 6, Determining the “Dial Factors”

The “dial factors” are defined as the number by which the difference in the observed photopeak channel from the standard photopeak channel is divided to obtain the dial setting change. This value, the dial setting change, is added to the old dial setting to obtain the new dial setting. The general method is performed in several steps:

  • set the dial at a known setting,
  • collect data with a source in front of the detector(s) until you have a statistically valid photopeak,
  • determine the center channel of the photopeak,
  • then repeat the previous three steps until you have sufficient data to determine the “dial factors.”
The balance of this chapter discusses the step-by-step method of calculating the “dial factors.”

11.6.1. Gain Setting

The gain is set at a nominal 5 keV/channel for all Helgeson counting systems. This does not mean that other gain settings could not be used, however, for most work with mixed fission, activation, and corrosion products, this is a convenient gain. If you are working mainly with natural thorium and its daughters you may wish to re-calibrate your equipment to use a gain setting of about 6 keV/channel since this will allow the 2.615 MeV photopeak of 208-thallium to be visualized completely.

11.6.2. Sources

Most of the time you will use the standard 60-cobalt sources, although you may wish to use 40-potassium in some form since no license from the Nuclear Regulatory Commission or any state regulatory body is required. The typical 0.5 to 1.0 microcurie 60-cobalt source will generally give a sufficient number of gamma rays at 1.1732 MeV to allow a calibration to be made in 30 to 60 seconds.

The trade-marked products “Lite Salt” or “No Salt,” contain a mixture of ordinary table salt (NaCl) and potassium chloride (KCl). An 11 ounce (311 gram) package of “Lite Salt” contains “550 mg sodium and 73 mg potassium...in each one-half teaspoon.” These containers have typical dimensions of 2.5 inches diameter by 5.375 inches high (approximately 65 by 135 mm). Thus, about six packages of these containers stacked in a postal mailing tube provides a column of 40-potassium which may be placed in from of the detectors, allowing simultaneous calibration of all detectors. This type of source will give a sufficient number of gamma rays at 1.4607 MeV to allow a calibration to be made in one to two minutes.

Thorium gas mantles may also be used since they provide a multitude of gamma rays. As in the case of the 40-potassium, they may be mounted on a column device so each detector views a source. One envelope, which contains two mantles, mounted in front of each detector will generally give a sufficient number of gamma rays at 2.615 MeV to allow a calibration to be made in one to two minutes.

Regardless of which sources you use the establishment of the gain factors is done in exactly the same manner.

11.6.3. Data Acquisition

There is no programmed method for determining the dial calibration factors since they are determined so infrequently. The best way to accumulate the data is to do it in Background Mode.

Let us assume that the detector, amplifier, and high voltage settings are approximately correct. This means that the photopeaks should be found near the following channels:

        Approximate
    Nuclide    Channel    Energy, keV    Exact Channel

    241-Americium    14    59.537    14.269

    60-Cobalt    230    1173.200    230.408

    60-Cobalt    261    1332.500    260.510

    40-Potassium    285    1460.800    284.670 11.6.4. Data Accumulation

a. Place the source(s) in front of the detector(s).

b. Start data accumulation by typing “ALT-R.”

c. Find the Center of the Photopeak

After about 30 seconds, or when there is a good photopeak, move the variable cursor to the apparent center of the photopeak, press “Page Up” to mark the photopeak, then press “P” (for peak information) and “1” (for marker 1) and read the center channel in the center of the screen, assume it is channel 234.5 for this example.

Record the dial setting and the center channel.

d. Repeat the previous four steps at about 5 or 6 different dial settings, where the first dial setting is that in which the center channel is closest to the ideal channel and the other five or six consist of two dial settings which are approximately 0.5 and 1.0 dial turns below the ideal channel and two dial settings which are approximately 0.5 and 1.0 dial turns above the ideal channel.
 
 
 

Typical Variation of 60-Cobalt Center Channel vs. Dial Setting,
Amplifier with the “Reciprocal” Gain vs Channel Curve

Dial Setting    Detector 1    Detector 2    Detector 3    Detector 4

    3.00                189.30            198.28            188.79            189.26

    3.50                198.36            201.47            197.65            198.30

    4.00                208.49            211.68            207.87            208.09

    4.50                220.27            223.41            219.52            220.12

    4.75                226.97            230.08            226.19            226.69

    5.00                233.66            236.75            232.85            233.25

    5.25                241.52            244.96            240.72            241.07

    5.50                249.79            253.81            249.18            249.44

    6.00                268.77            272.60            268.26            268.41

    6.50                 292.20            296.13            291.89            291.87

    7.00                320.62            325.30            320.59            320.48

 

Table 11-5, This is a table showing the change in the 60-Cobalt 1.17 MeV photopeak centers versus dial setting for four different NaI(Tl) detectors using the older non-linear
“reciprocal curve”  amplifier.

   

Table 11-5, above, and Figure 11-14, below, show the degree of non-linearity which you might expect from one of the older amplifiers. Note that this shows the change in the 60-cobalt 1.17 MeV center channel over 40 percent of the entire dial capacity, a condition which is not observed frequently, since most of the dial changes occur within 1 to 1.5 turns of the dial.
 

NONLNAMP.IMG
Figure 11-14, This graph shows the data of Table 11-5.

Now let us look at the linear amplifiers used in the newer Helgeson equipment.


Typical Variation of 60-Cobalt Center Channel vs. Dial Setting,
Amplifier with the “Linear” Gain vs Channel Curve

Dial Setting    Measured    Calculated    Difference, %

    1.00                 113.50           115.58           -1.837

    2.00                143.90            144.37           -0.330

    3.00                173.80            173.16            0.366

    4.00                203.60            201.95            0.808

    5.00                232.30            230.74            0.670

    6.00                260.30            259.53            0.294

    7.00                288.10            288.32           -0.078

    8.00                317.00            317.11           -0.036

    9.00                344.20            345.90           -0.495


Table 11-6, This is a table showing the change in the 60-Cobalt 1.17 MeV photopeak centers versus dial setting for one NaI(Tl) detectors using the newer Helgeson linear amplifier.

Table 11-6, above, and Figure 11-15, below, show the excellent linearity which you will observe from one of our newer amplifiers. Note that we cannot easily distinguish the difference between the plot of the original data and the plot of the calculated data. A correlation coefficient of 0.999863 explains why these values are so good.
 

LINAMP.IMG
Figure 11-14, This graph shows the data of Table 11-5.

11.6.5. Calculate the New Dial Factor

The calculations for determining the new dial factor may be made by several simple methods:

  • manually by simple subtraction and division; this method does not give an estimate of the statistical of the error of the measurements, or by
  • least squares, which is somewhat more complicated, but which gives an estimate of the statistical error of the measurements. This method may be done manually or using the “Curvefit” program which is discussed in Chapter of the “HELGE” manual.
11.6.5.1. Dial Factors by Simple Subtraction and Division

11.6.5.1.1 Example from a “Reciprocal” Gain vs Channel Amplifier

Let us use the data of Table 11-5 for our first example. We can see from the graph that the curve is not linear, which means that we must choose the range within which we will make our calculations. If we use dial settings from 4.5 through 5.5, we will have five data sets. This should be sufficient to determine the correct dial factor. Since the method is the important part to learn, we shall use only the data from Detector 1 shown below in Table 11-7.


Typical Calculation of the Dial Factor for an
Amplifier with the “Reciprocal” Gain vs Channel Curve

                                               Delta      Delta        Dial
 Dial Setting    Measured     Dial    Channel    Factor

    3.00                189.30

    3.50                198.36

    4.00                208.49

    4.50                220.27

    4.75                226.96            0.25       6.69       26.76

    5.00                233.66            0.25       6.70       26.80

    5.25                241.52            0.25       7.86       31.44

    5.50                249.79            0.25       8.27    33.08

    6.00                268.77

    6.50                292.20        Average Factor =    29.52

    7.00    320.62

Table 11-7 This is a table showing the change in the 60-Cobalt 1.17 MeV photopeak centers versus dial setting for four different NaI(Tl) detectors using the older non-linear
“reciprocal curve”  amplifier.

The third and fourth columns contain the differences between the dial settings and the center channels, respectively. The fifth (far right) column is obtained by dividing the channel differences by their corresponding dial differences. The bottom number in the fifth column is the arithmetic average of the four estimates of the dial factor. Therefore, you may now go to the “Examine/Change Parameters” selection from the Main Menu, select option 8, “Calibration,” and enter the correct dial factor for the particular detector you have just calibrated.
 

11.6.5.1.2. Example from a “Linear” Gain vs Channel Amplifier

Let us use the data of Table 11-6 and Figure 11-15 for our second example. We can see from the graph that the curve is very linear, which means that we may use all of the data for making our calculations. These calculations are shown in Table 11-8, below.

The third and fourth columns contain the differences between the dial settings and the center channels, respectively. The fifth (far right) column is obtained by dividing the channel differences by their corresponding dial differences. The bottom number in the fifth column is the arithmetic average of the four estimates of the dial factor. Therefore, you may now go to the “Examine/Change Parameters” selection from the Main Menu, select option 8, “Calibration,” and enter the correct dial factor for the particular detector you have just calibrated.

Typical Calculation of the Dial Factor for an
Amplifier with the “Linear” Gain vs Channel Curve
Using only Addition, Subtraction, and Division

                                               Delta      Delta         Dial
 Dial Setting    Measured     Dial    Channel    Factor

     1.00                113.50

     2.00                143.90            1          30.40       30.40

     3.00                173.80            1          29.90       29.90

     4.00                203.60            1          29.80       29.80

     5.00                232.30            1          28.70       28.70

     6.00                260.30            1          28.00       28.00

     7.00                288.10            1          27.80       27.80

     8.00                317.00            1          28.90       28.90

     9.00                344.20            1          27.20       27.20

                        Average                                       28.84

Table 11-8, This is a table showing the change in the 60-Cobalt 1.17 MeV photopeak centers versus dial setting for four different NaI(Tl) detectors using the older non-linear “reciprocal curve”  amplifier.

 

11.6.5.2. Dial Factors by Least Squares Calculations

The Theory of Least Squares is presented fully in Chapter 36. Please refer to this chapter for a complete discussion of the operations presented here.
 
Typical Calculation of the Dial Factor for an Amplifier with the
“Reciprocal” Gain vs Channel Curve Using the Least Squares Principles

      Dial Setting    Measured

              x                       y            x^2                x*y              y*2

           4.50               220.27        20.2500      991.215    48,518.8729

          4.75               226.96         22.5625    1,078.060    51,510.8416

          5.00               233.66         25.0000    1,168.300    54,597.9956

          5.25               241.52         27.5625    1,267.980    58,331.9104

          5.50               249.79         30.2500    1,373.845    62,395.0441

n =        5

Sum(x)    25

Sum(y)    1,172.25

Sum(x^2)    125.6250

Sum(x*y)    5,879.41

Sum(y^2)    275,353.6646

_VP_EQN_6.GIF

_VP_EQN_7.GIF

_VP_EQN_8.GIF

_VP_EQN_9.GIF

    a=    87.24

    b=    29.44, This, the slope, is the dial factor

Table 11-9, This is a table showing the change in the 60-Cobalt 1.17 MeV photopeak centers versus dial setting for four different NaI(Tl) detectors using the older non-linear "reciprocal curve" amplifier.

Last Update: 1-May-2003